MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3505430BC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
09/01/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: possible rupture, possible capsular contracture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a primary breast augmentation surgery with implantation of a 430cc mentor memorygel breast implant prosthesis.Post-operatively, the patient experienced right breast soreness after exercising.She observed the right breast became flat and the nipple/areola were lower than on the left.During a physical exam, she was noted to have a smaller right breast which was firm and mildly constricted/flat in the superior pole.A diagnosis of possible right breast capsular contracture (no baker grade was provided) or a ruptured right implant was provided.As a result, the patient underwent bilateral exchange with mentor gel implants on (b)(6) 2025.During the surgery, a ruptured right implant was confirmed.
|
| |
|
Additional Manufacturer Narrative
|
|
On january 07, 2026, the mentor analysis lab received the suspect medical device for evaluation.On january 14, 2026, mentor completed an evaluation on the returned device.Device evaluation: mentor conducted visual inspection.Visual analysis of the returned device revealed that the breast implant was found to be ruptured.The evaluation determined that the possible cause of the rupture is the trauma experienced.Rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Trauma is a known inherent risk of gel-filled mammary prosthesis.Regarding the reported capsular contracture condition, there are not enough details to determine the factors that may have caused this condition.Capsular contracture in the patient¿s breast might be the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Further, capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.On january 19, 2026, mentor was provided with the baker grade rating.The patient was diagnosed with right breast baker grade iii capsular contracture.Manufacturer¿s reference number: (b)(4).
|
| |
|