| Catalog Number |
MN1413 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2025, a patient underwent an ultrasound guide breast biopsy procedure by using the magnum.It was reported that during the procedure, the device needle allegedly break.The procedure was completed using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records was performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional information was received, and the file was reassessed for reportability based on the follow-up information, which states the plastic part broken outside the patient and determined to be no longer reportable.Since an initial mdr was submitted, the file will remain assessed as a malfunction.B5, g3, h6 (component).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, a patient underwent an ultrasound-guided breast biopsy through normal density tissue using the magnum instrument.During the procedure, the hub of the device was found to be broken.The procedure was completed using another device.The procedure was completed using another device.There was no reported patient injury.
|
| |
|
Search Alerts/Recalls
|