PERFUSION SYSTEMS EOPA ARTERIAL CANNULA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO
|
Back to Search Results |
|
| Model Number |
77522 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/06/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure involving these eopa arterial cannulae, the customer attempted to use the first cannula, however, there was an alleged product issue where the white introducer kept pushing back during cannulation.As a result, the first cannula was replaced.The second cannula, allowed cannulation, however as the introducer was withdrawn, blood started to leak out as if the introducer had not formed a seal.The cannulae were not cooled or heated and there was no visible damage to the cannulae or packaging.There was minimal blood loss of approximately 20mls and no fluid or blood transfusion was required.The second cannula was used to complete the procedure.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: visual inspection shows no outward signs of any damage.The dilator od was measured using a keyence scope and the hemostasis cap id was measured using a plug gage.Dilator od - was measured to be 0.239 inches, specification, 06819 has the od to be 0.242 +0.002/-0.004 inches hemostasis cap id - was measured to be 0.230 inches, specification has the id to be 0.225 +/- 0.005 inches the dilator was inserted through the cap and out the tip of the cannula.A force was applied to the tip of the introducer with no pushing back observed.Reason for the return was not confirmed.Additional information h6.2 eval code method (fdm/annex b): this field was updated.Additional information h6.3 eval code result (fdr/annex c): this field was updated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|