|
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h11.The microsensor (id 826633) was returned for evaluation.Device history record (dhr) - no anomalies were identified that occurred during the manufacture or packaging of the device.Failure analysis - both the catheter and sensor were examined; the catheter showed no surface damage; all connectors were inspected and found to be in good condition.The dark blue cap was not fully tightened.After tightening blue cap, the fittings were inspected.A syringe filled with water was used to introduce fluid into the catheter, and no leaks were observed.The complaint could not be confirmed.Root cause analysis - the most likely cause for the reported incident is the cap not being tightened fully.
|