MAUDE Adverse Event Report: DEPUY IRELAND - 3015516266 VELYS ARRAY SET KNEE; ROBOTIC-ASSISTED SOLUTION - DEVICE ARRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 3015516266 VELYS ARRAY SET KNEE; ROBOTIC-ASSISTED SOLUTION - DEVICE ARRAY Back to Search Results
Catalog Number 451570011
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2025, the patient underwent vras tka for oa with the array in question.During the surgery, although the device array was connected, it came loose with light force.And they tried to change the array interface, but there were no improvements.The surgeon used another new array for the surgery without any problems.The surgery was completed successfully within 30 minutes surgical delay.There was no patient consequence related to the event.No further information is available.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A manufacturing record evaluation was performed for the finished device product code: 451570011 lot number: mk289929, and no non-conformances or manufacturing irregularities were identified.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: a manufacturing record evaluation was performed for the finished device product code: 451570011 lot number: mk289929, and no non-conformances or manufacturing irregularities were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VELYS ARRAY SET KNEE
Common Device Name
ROBOTIC-ASSISTED SOLUTION - DEVICE ARRAY
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
elba bello
700 orthpaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23957978
Report Number1818910-2026-00059
Device Sequence Number12168739
Product Code OLO
UDI-Device Identifier10603295520146
UDI-Public(01)10603295520146(17)280630(10)MK289929
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K202769
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number451570011
Device Lot NumberMK289929
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/15/2025
Supplement Date Received by Manufacturer03/12/2026
Initial Report FDA Received Date01/05/2026
Supplement Report FDA Received Date03/13/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/09/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-