Product complaint # (b)(4) date sent to the fda: 1/5/2026.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested; the following was obtained: if available, please provide the lot number of the products involved received information as following from sales rep via phone today.The lot number is unknown.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
|
|
Product complaint # (b)(4).Corrected information: d 9.Device available for evaluation, h 3.Device evaluated by manufacturer? additional information: d 9.Date device returned to manufacturer, d 9.Is device returned to manufacturer, h 6.Component code, h 6.Type of investigation, h 6.Investigation findings, h 6.Investigation conclusions.H3 investigation summary: the product was returned to eth for evaluation.One empty labeled winding former and one needle suture pieces that pertain to product code vcp358h were received for analysis.During visual inspection of the returned sample, marks that appear to be caused by a surgical instrument were observed on the body needle.The suture was examined, and the end was noted to be frayed.Besides that, it was cut probably caused by a surgical instrument.In addition, body fluids were found on the strand.The other section of the suture was not returned for analysis.Since the sample was received open, the functional test could not be performed due to undetermined exposure time to the environment.Care should be taken to avoid damage to the strand when handling and avoid the crushing or crimping action of surgical instruments such as needle holders and forceps.As part of eth quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.The manufacturing records could not be reviewed as the batch/lot number is unknown.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|