• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATUN PRESSFIT STR STEM14X110MM; KNEE STEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATUN PRESSFIT STR STEM14X110MM; KNEE STEMS Back to Search Results
Catalog Number 151314110
Medical Device Problem Codes Loss of or Failure to Bond (1068); Osseointegration Problem (3003)
Health Effect - Clinical Codes Inadequate Osseointegration (2646); Fibrosis (3167)
Date of Event 12/16/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient underwent the revision of an attune due to loose femoral and tibial components.The surgeon noted fibrous tissue up the femoral canal around the femoral stem causing loosening.The tibial component was loose as it appeared that the surgeon had cemented in the cementless tibial sleeve.Resulting in no bone ingrowth.The patella component was retained from previous primary tkr.The loosening was between the implant and bone interfaces.Affected side: right side.
 
Event or Problem Description
Additional information received states that loosening between the implant and bone interfaces.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary- no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot- the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H11 additional narrative: added: b5, d10 (concomitant), h6 (health effect - clinical code).Corrected: d4 (primary udi number, expiry date), h4, h6 (medical device problem code).Updated pec code from implant loosening interface - unknown to implant loosening interface - implant to bone, and patient code inadequate osteointegration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATUN PRESSFIT STR STEM14X110MM
Common Device Name
KNEE STEMS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthpaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23958190
Report Number1818910-2026-00065
Device Sequence Number5610330
Product Code JWH
UDI-Device Identifier10603295043614
UDI-Public(01)10603295043614(17)280101(10)HN9865
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number151314110
Device Lot NumberHN9865
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/15/2025
Supplement Date Received by Manufacturer01/09/2026
Initial Report FDA Received Date01/05/2026
Supplement Report FDA Received Date01/09/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CRS FEMORAL RT SZ 5 CEM.; ATTUNE POS FEM AUG SZ 5 8MM.; ATTUNE PS RP INSRT SZ 5 8MM.; ATUN TIB SLV M/L 37MM HALF POR.; ATUNE PRESSFIT STR STEM12X60MM.; ATUNE REV RP TIB BASE SZ 4 CEM.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-