Catalog Number 210133
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/17/2025
Type of Reportable Event Malfunction
Event or Problem Description
This is report 1 of 2 of (b)(4).It was reported that during an arthroscopy surgical procedure it was observed that the rigidfix fem 3.3mm s/t xpin device anchor was broken off.All broken parts were removed.Changed another one to continue the surgery, the same problem happened again.Another like device was used to complete the procedure.There was no reports of delay in the procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: d9, h6: the actual device has been returned and is currently pending evaluation.Once the investigation has been completed, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found the two sets of cross pins deformed.The overall complaint was not confirmed as the observed condition of the rigidfix fem 3.3mm s/t xpin would not contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to thermal degradation of the polymers due to an exposure to higher temperatures than the recommended while in stock.As per ifu, properly handling and attention to the approved use of the device diminishes the risk of failure.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.There was no non-conformance regarding this lot.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found the two sets of cross pins deformed.The overall complaint was not confirmed as the observed condition of the rigidfix fem 3.3mm s/t xpin would not contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to thermal degradation of the polymers due to an exposure to higher temperatures than the recommended while in stock.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RIGIDFIX FEM 3.3MM S/T XPIN
Common Device Name
PIN, FIXATION, SMOOTH
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
MEDOS INT MITEK
chemin blanc 38
le locle CH-02 400
SZ   CH-02400
Manufacturer Contact
elba bello
325 paramount drive
raynham, MA 02767
9083863534
MDR Report Key23958613
Report Number1221934-2026-00036
Device Sequence Number20236418
Product Code HTY
UDI-Device Identifier10886705000890
UDI-Public(01)10886705000890(10)214L071(17)260831
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K974341
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2026
Device Catalogue Number210133
Device Lot Number214L071
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/16/2025
Supplement Date Received by Manufacturer01/12/2026
01/16/2026
02/09/2026
Initial Report FDA Received Date01/05/2026
Supplement Report FDA Received Date01/15/2026
01/21/2026
02/10/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/15/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-