• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 GROSHONG NXT CLEARVUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 GROSHONG NXT CLEARVUE Back to Search Results
Catalog Number 7812400
Medical Device Problem Code Component Incompatible (1108)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Each event reported to bd is evaluated and investigated in accordance with our complaint investigation procedures.The investigation process includes, but is not limited to, evaluation of the event details provided by the complaint facility, a review of complaint history, manufacturing records and risk document where applicable, and an evaluation of the subject device when available to identify potential contributing factors.The device has not been returned to the manufacturer for evaluation.
 
Event or Problem Description
It was reported that the connection was lost during the installation process.It was reported this occurred with six devices.This report addresses all six devices.No other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GROSHONG NXT CLEARVUE
Common Device Name
GROSHONG NXT CLEARVUE
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
maddy vincent
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key23960624
Report Number3006260740-2025-10205
Device Sequence Number15855916
Product Code LJS
UDI-Device Identifier00801741075261
UDI-Public(01)00801741075261(11)241128(17)290930(10)REJX1638
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K034020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/31/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number7812400
Device Lot NumberREJX1638
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/16/2025
Initial Report FDA Received Date01/05/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/28/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-