ZIMMER BIOMET, INC. TPRLC XR T1 PPS 15X150MM; PROSTHESIS, HIPS Back to Search Results
Model Number N/A
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).G2: foreign, event occurred in japan.The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that during the inspection of circulated stock, it was identified that the sterile packaging was damaged.There was no patient involvement.
 
Event or Problem Description
No additional information available for the reported event.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: d9; g3; h2; h3; h6.Visual evaluation of the returned product/provided photos identified damage to the sterile packaging (pouch).Sterility has been compromised.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The root cause of the reported event can be attributed to transit damage and a packaging design issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC XR T1 PPS 15X150MM
Common Device Name
PROSTHESIS, HIPS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23961321
Report Number0001825034-2026-00037
Device Sequence Number13143748
Product Code LPH
UDI-Device Identifier00880304517165
UDI-Public(01)00880304517165(17)320316(10)7207504
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K120030
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberN/A
Device Catalogue Number51-105150
Device Lot Number7207504
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/08/2025
Supplement Date Received by Manufacturer05/28/2026
Initial Report FDA Received Date01/05/2026
Supplement Report FDA Received Date06/08/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-