DEPUY MITEK LLC US MILAGRO ADVANCE SCREW 8X23MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
231817 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unspecified surgical procedure of the knee joint it was observed that the milagro advance screw 8x23mm device broke and 2x 5.5 healix advance kntlss br anchor devices broke.It was reported that all broken pieces were removed from the patient.Another device was used to complete the procedure.There were no adverse consequences to the patient.No additional information could be provided.This is report 1 of 3 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: d9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative:
investigation summary: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found that the screw was broken near mid-body.Not all the broken pieces returned for evaluation.Additionally, the device had foreign matter, presumably biological matter.No other anomalies were noted.The overall complaint was confirmed as the observed condition of the milagro advance screw 8x23mm would have contributed to the complained device issue.Based on the investigation findings, the potential cause is traced to the procedural variables, such handling of the device or product interaction during procedure.Resistance may have been felt when the device was being inserted and the excessive force applied caused the screw to break.As per the instructions for use (ifu): 1) if resistance is felt during the insertion of a milagro br interference screw over a guidewire, stop and confirm that the guidewire is not entrapped.If entrapped, back out the screw and withdraw the guidewire.Properly handling and attention to the approved use of the device diminishes the risk of failure.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
| |
|