| Model Number |
RLM274K |
| Medical Device Problem Code |
Erratic or Intermittent Display (1182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A steris service technician arrived onsite following the event to inspect the monitor and was unable to duplicate the reported event.The surgical monitor subject of the reported event was returned to steris service depot for evaluation, and the issue was unable to be reproduced.The user facility was provided with a replacement monitor.No additional issues have been reported.
|
| |
|
Event or Problem Description
|
|
The user facility reported that during a patient procedure the image to their vividimage 4k surgical monitor loses focus and intermittently displays black lines.The procedure was completed successfully with no report of injury.
|
| |
|