The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The device was returned to a third-party service center for evaluation.During the evaluation of the device, the service technician observed visible foam particles inside the blower kit and a blower foam degradation.Additionally, the service technician also noted dust fail contamination.There was also error codes present in the device logs: e066 (err_stuck_encoder_b).The device was scrapped.
|