| Catalog Number |
367815 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported prior to using bd vacutainer® serum blood collection tubes, abnormal additive was seen with one tube.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd received a sample and a photo for investigation.The sample was visually inspected for additive abnormality and failed inspection.The photos were reviewed and the indicated failure mode for additive abnormality was observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode: additive abnormality.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported prior to using bd vacutainer® serum blood collection tubes, abnormal additive was seen with one tube.No adverse impact was reported regarding the patient or user.
|
| |
|
Search Alerts/Recalls
|