• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMVIE US CORP LLC OSSEOTITE® TAPERED CERTAIN® PREVAIL® IMPLANT 6/5 X 10MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMVIE US CORP LLC OSSEOTITE® TAPERED CERTAIN® PREVAIL® IMPLANT 6/5 X 10MM; DENTAL IMPLANT Back to Search Results
Catalog Number XIITP6510
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Erosion (1750); Bone Fracture(s) (1870)
Date of Event 12/05/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A4: weight unknown / not provided.Product has been received by zimvie and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the implant at tooth site #18 was removed due to lack of primary stability associated with interseptal bone fracture.An additional appointment was anticipated to replace the implant.Symptoms as a result of the event: abscess.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation code was added: 10, 4109, 3331 and 4111.H6: investigation findings code was added: 213.H6: investigation conclusions code was added: 4315.H10: narrative/data was updated.Zimvie received the reported device for evaluation.Visual evaluation of the as returned device identified the implant with signs of use.However, no apparent damage/malfunction to the implant was identified that would cause or contribute to the reported events.Markings are likely due to the removal process.Functional testing to recreate the reported events could not be performed due to the nature of the device and events.This complaint refers to the specific device being investigated for this complaint record.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported events, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.The customer did not submit images for the reported events.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are patient condition/patient factors.Therefore, based on the available information, a device malfunction did not occur.The reported events were non-verifiable with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.Similar complaints for the lack of primary stability (unable to place implant into osteotomy) have been previously investigated.Visual and dimensional evaluations of the previously returned product have not identified or suggested manufacturing non-conformances.While non-conformances were identified for some lots during manufacturing records reviews, the documented disposition actions for each have not suggested the likely release of non-conforming product.To date, all complaint data was found to be conforming and did not meet capa/hhe/d/ or any further escalations.Therefore, there were no complaints which confirmed a manufacturing or design related issue caused or contributed to the reported event.
 
Event or Problem Description
No additional event information received at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE® TAPERED CERTAIN® PREVAIL® IMPLANT 6/5 X 10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark  mashburn
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key23968044
Report Number0001038806-2026-00042
Device Sequence Number4928968
Product Code DZE
UDI-Device Identifier00844868022206
UDI-Public(01)00844868022206(17)290826(10)2120017570
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K130949
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 02/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Catalogue NumberXIITP6510
Device Lot Number2120017570
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer12/12/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/05/2026
Supplement Report FDA Received Date02/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/27/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexMale
Patient EthnicityNon Hispanic
-
-