• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. 2.7 DEGREE STRAIGHT SAGITAL SAW; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. 2.7 DEGREE STRAIGHT SAGITAL SAW; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 212186
Medical Device Problem Codes Mechanical Problem (1384); Positioning Problem (3009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/09/2025
Type of Reportable Event Malfunction
Event or Problem Description
The following information was provided: sawblade was staggering and vibrating loudly.Case type / application: tka 2.0.
 
Additional Manufacturer Narrative
An event regarding mechanical failure involving a mako saw attachment was reported.The event was not confirmed because the product was not available for inspection.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been other complaints with similar event(s) for the lot referenced.Conclusions: the alleged failure mode was not confirmed because the product was not available for inspection.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
 
Event or Problem Description
No new information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7 DEGREE STRAIGHT SAGITAL SAW
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAKO SURGICAL CORP.
3365 enterprise ave
weston FL 33331
Manufacturer (Section G)
MAKO SURGICAL CORP.
3365 enterprise ave
weston FL 33331
Manufacturer Contact
anna ryan
raheen business park
limerick, FL NA
EI   NA
9546280700
MDR Report Key23968223
Report Number3005985723-2026-00006
Device Sequence Number17330100
Product Code OLO
UDI-Device Identifier00848486032111
UDI-Public00848486032111
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K172219
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number212186
Device Lot Number35220924
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/09/2025
Supplement Date Received by Manufacturer01/16/2026
Initial Report FDA Received Date01/05/2026
Supplement Report FDA Received Date02/12/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/26/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-