• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN AG JUAREZ (MMJ) SHILEY; TUBE, TRACHEAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN AG JUAREZ (MMJ) SHILEY; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number 18885
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the endotracheal tubes top adapter can easily be remove from the tube itself with minimal effort while still in the package.There was no patient involved.
 
Additional Manufacturer Narrative
D10 concomitant product: 18885, 18885 8.5mm taperguard evac trachealx10, lot# 25g0377jzx.18880, 18880 8.0mm taperguard evac trachealx10, lot# 25f0326jzx.18875, 18875 7.5mm taperguard evac trachealx10, lot# 25g0050jzx.18875, 18875 7.5mm taperguard evac trachealx10, lot# 25i0307jzx.86450, 86450-w 7.0mm inter hi-lo trach x10, lot# 25e0044jzx.18870, 18870 7.0mm taperguard evac trachealx10, lot# 25h0038jzx.18865, 18865 6.5mm taperguard evac trachealx10, lot# 25g0392jzx.18865, 18865 6.5mm taperguard evac trachealx10, lot# 25g0389jzx.18860, 18860 6.0mm taperguard evac trachaelx10, lot# 25g0403jzx.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3 and h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the endotracheal tubes top adapter can easily be remove from the tube itself with minimal effort while still in the package.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX  32575
Manufacturer (Section G)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX   32575
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key23968541
Report Number9681384-2026-00015
Device Sequence Number4928980
Product Code BTR
UDI-Device Identifier40884521032553
UDI-Public(01)40884521032553(17)300704(10)25G0376JZX
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K090352
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number18885
Device Catalogue Number18885
Device Lot Number25G0376JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/05/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/10/2025
Supplement Date Received by Manufacturer02/17/2026
Initial Report FDA Received Date01/05/2026
Supplement Report FDA Received Date02/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/04/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NOTES ON H11
Patient SexUnknown
-
-