|
Additional information provided in d.9., h.3., h.6.And h.11.The used infusion manifold was visually inspected, and no obvious defects were found.The returned infusion manifold was found to be in good condition.One lab stock pak was used to test the returned product.The infusion manifold connectors were connected to the cassette without any interference.A calibrated console representing the current software version was used to test the product.The product could prime and pass intraocular pressure (iop) calibration successfully.No anomalies were observed during priming.The infusion pressure was measured at multiple set points throughout the console range and met specifications.No system message code was generated during testing.Toggling the infusion and the fluid/air exchange (f/ax) modes, fluid and air flowed from the cassette to the infusion line continuously without any bubble in various settings in all sub modes.With the infusion control on, fluid flowed from the cassette continuously without generating air to the infusion line.When the fluid/air exchange (f/ax) control was on, only air was introduced from the cassette to the infusion line.Fluid flowed from the balanced salt solution (bss) bottle to the drain bag without any interfering.No leakage was detected from the pump elastomer or on the pump area of the fluidics module.Cleaning process was able to be performed after functional test had completed.We were unable to replicate the customer's experience during laboratory evaluation.Fluid flowed through the fluidics management system as it should during functional testing.No anomalies were observed.The investigation conducted a non-conformance review of the reported lot number.No deviation was identified that would have contributed to this event and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event as the product functioned per specifications.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint, therefore no action will be taken for this occurrence.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|