• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COMFORT; UNO COMFORT 110/17 SC1 MINI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COMFORT; UNO COMFORT 110/17 SC1 MINI Back to Search Results
Model Number MMT-377A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 11/12/2025
Type of Reportable Event Malfunction
Event or Problem Description
Reference number (b)(4).Event occurred in the united states.It was reported that patient faced infusion set tubing detachment event on (b)(6) 2025.The insertion site was the left abdomen.The tubing got detached from connector.Infusion set was used for atleast two days.Blood glucose level was high at the time of the event.No further information available.
 
Additional Manufacturer Narrative
E1: patient city: (b)(6).Patient country: united states.
 
Additional Manufacturer Narrative
Correction: this mdr is being submitted to correct the submitted expiration date under d4.Additional information - this mdr is being submitted to include the below: h6: investigation results under type of investigation, investigation findings, investigation conclusions.H11: investigation summary.Complaint investigation results: electronic quality management system (eqms) search: a query was run in the eqms on 20/apr/2026 against "lot number" "6012518" and similar malfunction codes: leak between tubing and site connector detachment / significant wetness, leakage from connection between the tubing connector and the infusion set (cannula part/base piece), tubing detached from tubing-tubing connector.The review confirmed that lot 6012518 and the identified failure mode are not associated with any nonconforming reports (ncr)s or corrective and preventive action (capa)s of the same or similar nature.Similar complaints search: a query was run in the eqms on 20/apr/2026 against "lot number" criteria equal "6012518" and similar malfunction codes: leak between tubing and site connector detachment / significant wetness, leakage from connection between the tubing connector and the infusion set (cannula part/base piece), tubing detached from tubing-tubing connector.The count of complaints is 1.The complaints numbers are 2457346.Device history record (dhr) review: the lot 6012518 was manufactured according to work instruction (wi) version 21 and manufactured in the machine multivac 14, on 01/apr/2025 with a total of (b)(6) units.Glue-tubing lot: the lot 5b04786 was manufactured according to wi version 15 and manufactured in the machine lc01, on 24/feb/2025 with a total of (b)(6) units.The lot 5c04053 was manufactured according to wi version 17 and manufactured in the machine lc01, on 26/mar/2025 with a total of (b)(6) units.The device history record (dhr) was reviewed in accordance with applicable procedures.All required in-process and final tests were completed and met specified requirements.No deviations were identified, and no maintenance events were recorded that relate to the complaint code.Conclusion: dhr review identified minor findings.They were managed according to procedure and did not impact compliance; no issues noted.Visual evidence review: no photo was provided to support visual confirmation of the reported issue.Conclusion: unable to perform visual verification; assessment based on available documentation only.Capa determination assessment - conclusion summary of complaint investigation: based on the investigation, no further investigation is required.The record will be closed and monitored through tracking and trending per wi (monthly trips and alerts).Complaint unconfirmed: following investigation, the reported failure could not be confirmed for this complaint.Conclusion summary of complaint investigation: as a result of the following a comprehensive review was conducted, including eqms queries, similar complaint searches, device history record review, visual evidence assessment, and capa determination.No ncrs or capas of the same or similar nature were found for lot 6012518 and related malfunction codes for leak issues.One complaint was identified for this lot, but no trend or systemic issue was detected.The manufacturing records confirmed that the lot was produced in compliance with all requirements, with no deviations or maintenance events noted.No photo evidence was provided, so the assessment was based on documentation only.Based on these results, no manufacturing or quality issues were identified, and no further investigation is required.The record will be closed and monitored through routine tracking and trending.
 
Event or Problem Description
To date no additional patient or event details have been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMFORT
Common Device Name
UNO COMFORT 110/17 SC1 MINI
Manufacturer Contact
aaholmvej 1-3
osted
lejre, 4320
MDR Report Key23971364
Report Number3003442380-2025-17796
Device Sequence Number15710831
Product Code FPA
UDI-Device Identifier05705244019522
UDI-Public05705244019522
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K162812
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-377A
Device Lot Number6012518
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/31/2025
Supplement Date Received by Manufacturer04/20/2026
Initial Report FDA Received Date01/06/2026
Supplement Report FDA Received Date04/26/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-