MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC. METRX II; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC. METRX II; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 9560101
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/10/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: evaluation of the returned device was not completed at the time of this report.A follow up report will be sent when analysis is complete.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Information was received from a manufacturing representative regarding a scope.It was reported that there is image abnormality as lens is not clear.There was no patient involvement.There were no further complications reported regarding the event.
 
Additional Manufacturer Narrative
H3: product analysis :product:9560101, lotno:1177121 the reported condition/complaint (lens is not clear) was duplicated and it was cause by bent outer tube.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METRX II
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC.
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC.
4340 swinea rd
memphis TN 38118
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key23972011
Report Number3003120897-2026-00029
Device Sequence Number18465433
Product Code LXH
UDI-Device Identifier00673978085456
UDI-Public(01)00673978085456(11)170208(10)1177121
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K993021
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number9560101
Device Catalogue Number9560101
Device Lot Number1177121
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/10/2025
Supplement Date Received by Manufacturer03/09/2026
Initial Report FDA Received Date01/06/2026
Supplement Report FDA Received Date03/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/08/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-