| Model Number |
DVPB3D4 |
| Medical Device Problem Codes |
Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
|
| Health Effect - Clinical Codes |
Pain (1994); Dizziness (2194); Confusion/ Disorientation (2553); Shock from Patient Lead(s) (3162); Presyncope (4410)
|
| Date of Event |
12/12/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient received anti-tachycardia pacing (atp) and high voltage therapy from the single chamber implantable cardioverter defibrillator (icd) for possible atrial arrhythmia with rapid ventricular response that the device classified as ventricular fibrillation (vf).The patient experienced dizziness and/or presyncope.The icd remains in use. no further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was further reported via the following clinic that the patient was hospitalized, and it was confirmed that the icd delivered inappropriate shock for atrial fibrillation (af) with rapid ventricular response.The patient also experienced pain and confusion.The icd was reprogrammed, and the patient¿s medication was adjusted.
|
| |
|