| Model Number |
S-45-120-120-P6 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
It was reported that during preparation of the 4.50x120mm supera self expanding stent system (ses), it was noted the tip was separated; therefore, the device was not used.There was no patient involvement and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for analysis.The reported tip material separation was able to be confirmed.Production record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint handling database identified no other incidents from this lot.It is possible that inadvertent mishandling during removal from the packaging tray and/or during preparation for use resulted in the reported/noted tip material separation; however, this cannot be confirmed.The investigation determined a conclusive cause for the reported/noted difficulties cannot be determined.Based on the reported information and results of the complaint investigation there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that during preparation of the 4.50x120mm supera self expanding stent system (ses), it was noted the tip was separated; therefore, the device was not used.There was no patient involvement and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|