| Catalog Number |
366594 |
| Medical Device Problem Code |
Difficult or Delayed Activation (2577)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturing location for this product is (htl).This site is an oem manufacturing site.Therefore, bd corporate headquarters in franklin lakes, nj has been listed in sections d.3.And g.1.And the franklin lakes fda registration number has been used for the manufacture report number.E1: (b)(6).A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that when using one (1) bd microtainer® contact-activated lancet, the unit failed to retract after pressing the button.No adverse impact was reported regarding the patient or user.
|
| |
|
Event or Problem Description
|
|
It was reported that when using one (1) bd microtainer® contact-activated lancet, the unit failed to retract after pressing the button.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d9: device available for evaluation: yes.H.3 device eval by manufacturer? yes.D9: returned to manufacturer on: 02-jan-2026.Investigation summary: bd received 1 sample for investigation.The samples were evaluated by visual examination and the indicated failure mode for does not retract with the incident lot was not observed.Additionally, retention samples from bd inventory were evaluated by functional testing and no issues were observed relating to does not retract as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode does not retract.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|