| Catalog Number |
IAB-S840C |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/04/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported "after the catheter inserted into the patient, the balloon was found leakage.Change the new catheter".No patient injury or consequene reported.The patient's current condition is reported as "fine".
|
| |
|
Event or Problem Description
|
|
It was reported "after the catheter inserted into the patient, the balloon was found leakage.Change the new catheter".No patient injury or consequene reported.The patient's current condition is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).The serial number on the returned sample is (b)(4).Returned for investigation was a 40cc 8.0fr rediguard intra-aortic balloon catheter (iabc) without the original packaging.The sample was returned in a cardboard box and was in a ziploc bag.Upon return, the one-way valve was tethered to the short driveline tubing.The bladder was fully unwrapped.Dried blood was noted on the exterior surfaces of the returned sample.Some blood was noted within the helium pathway.The bladder thickness was measured at six points with measurements ranging from 0.0060in-0.0066in and was within specification of process document.The one-way valve was tested and passed.A vacuum was pulled on the one-way valve, and it held for at least 1 minute according to quality system document.The catheter's central lumen was aspirated and flushed using a 60cc lab-inventory syringe.No abnormalities or debris were noted.The iabc was leak tested in accordance with testing methods from manufacturing procedure.A leak was immediately detected from the iabc bladder membrane.Under the microscopic inspection, the leak site was consistent with contact from a sharp object and was noted at approximately 25.1cm from the iabc distal tip.No other leaks were detected.A lab inventory 0.025in guidewire was back loaded through iabc distal tip.No resistance was noted; the guidewire was able to advance through the central lumen.No blood or debris was noted.A device his tory record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint for iabc "balloon was found leakage" is confirmed.During the investigation, a puncture consistent with contact from a sharp object was found on the iabc bladder and this caused the reported complaint.No other leaks were detected during functional testing.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the bladder leak.The probable root cause is customer handling related.No further action is required at this time.This will be monitored for any developing trends.
|
| |
|
Search Alerts/Recalls
|