| Model Number |
P8005A |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 09-dec-2025, a other health care professional contacted technical service to report that transtar stretcher (product code p8005a000002, serial number (b)(6)), had the brake not holding.The process step during which this occurred was unknown.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
No further information is available on the repair of the bed at this time.The investigation is ongoing.However, if any additional relevant information is identified following completion of the investigation, the additional relevant information will be submitted in a supplemental report.
|
| |
|
Additional Manufacturer Narrative
|
|
The baxter technician found the casters needed to be replaced.A search of the baxter maintenance records did not show baxter performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.Per the baxter service manual the bed should be subject to an effective maintenance program.An annual service of the bed is advised in order to maintain its characteristics and performance.Brake casters should be checked for cuts, wear and quality of tread, etc.And replaced when necessary.Apply the brake and make sure the stretcher does not move.If there is movement, look at the brake components for wear.Apply the steer, and make sure the stretcher steers correctly.Look at the steer components for wear.Put the stretcher in neutral.Make sure all four casters rotate and roll freely.Adjust or replace components if necessary.The technician replaced the casters to resolve the reported event.Based on this information, no further action is required.
|
| |
|
Search Alerts/Recalls
|