| Model Number |
620-0211511-01 |
| Medical Device Problem Codes |
Pacing Problem (1439); Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device displayed a "pace defib device failure" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.During the investigation the technician discovered that the customer attempted repair of the device.As a result of the device being tampered with, root cause could not be firmly established.It was recommend to the customer to replace the processor/bridge/pace board as a precaution.Analysis of reports of this type has not identified an increase in trend.
|
| |
|