DEPUY SYNTHES PRODUCTS LLC 6.5CM ADULT CRANI ATTACHMT; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE
|
Back to Search Results |
|
| Catalog Number |
CRANI-A |
| Medical Device Problem Codes |
Device Damaged by Another Device (2915); Difficult or Delayed Separation (4044)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the neuro tip of the craniotome device was bent/damaged, the device had a worn bearing, and was difficult to assemble/disassemble.It was noted that there was a dura guard wound, the motor connection was snagged, and the bar was snagged.It was further determined that the device failed pretest for visual assessment, lock operation assessment, and cutter insertion assessment.It was noted in the service order that it was difficult to remove the cutter from the device.This event did not occur during surgery.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported; however, it was reported that the event occurred in 2025.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause of this condition was determined to be traced to component failure due to wear.
|
| |
|
Search Alerts/Recalls
|
|
|