DEPUY ORTHOPAEDICS INC US ALTRX +4 10D 32IDX50OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Catalog Number 122132150
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: h3, h6: no device associated with this report was received for examination.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient underwent tha with pinnacle and c-stem amt in question on an unknown date.Primary surgery was completed successfully without any surgical delay.The patient experienced loosening of artificial hip joint and gluteal hip dislocation, and penetration of the acetabular side of the product into the inner plate.The second surgery was performed on (b)(6) 2024.During removal, the cup, screw, and liner were removed as a single unit, and the tip of the screw broke off and remained inside the body.However, the surgeon decided that it would be difficult to remove, so he left the broken tip in place.((b)(4) was newly created to capture the event f broken screw tip) the head was replaced with a deltats36+12, and the surgery was completed.The surgeon commented that the cup placement angle during the primary surgery was poor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTRX +4 10D 32IDX50OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthpaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23975715
Report Number1818910-2026-00193
Device Sequence Number13418418
Product Code LPH
UDI-Device Identifier10603295016168
UDI-Public(01)10603295016168(17)260831(10)JF1719
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K072963
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2026
Device Catalogue Number122132150
Device Lot NumberJF1719
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/16/2025
Initial Report FDA Received Date01/06/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient Weight55 KG
-
-