| Catalog Number |
367921 |
| Medical Device Problem Code |
Short Fill (1575)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® sodium fluoride/potassium oxalate 5mg/4mg, an unspecified number of tubes underfilled.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reporter facility name: (b)(6).A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® sodium fluoride/potassium oxalate 5mg/4mg, an unspecified number of tubes underfilled.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary bd received a photo for investigation for catalog 367921, lot number 5048336, which also shows a shelf pack of samples without the customer¿s indicated failure mode.Functional tests were performed on 10 retained samples, and all tubes were within specification limits.For retained samples from lot 5048336, 100 samples were visually inspected with the hemogard closure assembly correctly assembled and placed on the tubes.Functional tests were performed on 10 retained samples, and all tubes were within specification limits.Based on a review of the device history record for the incident lots, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the lot 5048336 for the indicated failure mode: draw volume.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Search Alerts/Recalls
|