HEARTSINE TECHNOLOGIES LTD PACKAGE,350P,PP01,EN,350-STR-US-10; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 350P
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/12/2025
Type of Reportable Event Malfunction
Event or Problem Description
The customer contacted heartsine to report their device would not power on.In this state the device may not be able to deliver defibrillation therapy if needed.There was no patient involvement reported with the event.
 
Additional Manufacturer Narrative
Heartsine has requested the return of the device for investigation.Upon completion, the conclusions will be submitted in a follow-up report.
 
Event or Problem Description
The customer contacted heartsine to report their device would not power on.In this state the device may not be able to deliver defibrillation therapy if needed.There was no patient involvement reported with the event.
 
Additional Manufacturer Narrative
A service representative performed evaluation of the customer¿s device and was able to verify the reported issue.Upon receipt the device failed to power on upon pressing the on/off button.The fault was attributed to corrosion on the on/off button and its contact pads on the membrane pcb.This had resulted in the on/off button dome failing to make contact with its contact pads on the membrane panel; therefore, the device was unable to be powered on.Furthermore, corrosion on the underside of the shock button dome had prevented the shock button from functioning.The corrosion indicates that the device was stored outside of the indicated conditions.The device shall be scrapped and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACKAGE,350P,PP01,EN,350-STR-US-10
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
HEARTSINE TECHNOLOGIES LTD
207 airport road west
belfast BT3 9 ED
EI  BT3 9ED
Manufacturer (Section G)
HEARTSINE TECHNOLOGIES LTD
207 airport road west
belfast BT3 9 ED
EI   BT3 9ED
Manufacturer Contact
brian blakeslee
207 airport road west
belfast BT3 9-ED
EI   BT3 9ED
4258674577
MDR Report Key23976921
Report Number3004123209-2026-00021
Device Sequence Number18796904
Product Code MKJ
UDI-Device Identifier05060167120671
UDI-Public05060167120671
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
P160008
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number350P
Device Catalogue Number350-STR-US-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/12/2025
Supplement Date Received by Manufacturer04/17/2026
Initial Report FDA Received Date01/06/2026
Supplement Report FDA Received Date04/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/26/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-