• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH POLY SCREW, FEN, 6MM X 45; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH POLY SCREW, FEN, 6MM X 45; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 5560126045F
Medical Device Problem Codes Device Slipped (1584); Migration (4003)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/16/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint#: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11: additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Event or Problem Description
It was reported the patient underwent a scoliosis surgery on (b)(6) 2025.Post-surgery it was discovered the set screws had loosened so a revision procedure would be necessary on (b)(6) 2025.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: b5.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Additional information received: a.Was there any patient consequences or not ? - yes, as mentioned a revision surgery was needed to insert new trialtis innies.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: h3, h6: an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Based on the information available, there is no indication that a design or manufacturing issue has caused the reported complaint condition, therefore it has been determined that no corrective and preventive action is required at this time.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d4 (lot).Corrected: d1, d4 (catalog, primary udi number), h6 (medical device problem code).Removed pec functional loose.D4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h3, h6: an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Based on the information available, there is no indication that a design or manufacturing issue has caused the reported complaint condition, therefore it has been determined that no corrective and preventive action is required at this time.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.Device history review (dhr): lot # provided is not a valid number, therefore, the dhr could not be completed.If device is returned or lot number can be confirmed, the dhr will be revisited.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.H3, h4, h6: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found nothing indicative of a device nonconformance.No observations pertaining to the nature of the reported event could be identified as there was no x-ray evidence.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was not confirmed as the observed condition of the poly screw, fen, 6mm x 45 would have not contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a dhr evaluation was performed for the finished device product code: 5560126045fs-12 lot number: 433110 it was electronically reviewed and no nonconformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 24.06.2025.Manufacturing site: jabil le locle.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d9.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLY SCREW, FEN, 6MM X 45
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ   CH-2400
Manufacturer Contact
elba bello
325 paramount drive
raynham, MA 02767
9083863534
MDR Report Key23978097
Report Number1526439-2026-00004
Device Sequence Number4630023
Product Code NKB
UDI-Device Identifier10705034594415
UDI-Public(01)10705034594415(10)433110
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K233684
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 01/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number5560126045F
Device Lot Number433110
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/16/2025
Supplement Date Received by Manufacturer01/14/2026
01/21/2026
01/23/2026
02/10/2026
02/17/2026
03/12/2026
Initial Report FDA Received Date01/06/2026
Supplement Report FDA Received Date01/20/2026
01/22/2026
01/29/2026
02/15/2026
02/19/2026
03/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/24/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexFemale
-
-