MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL GALAXY G3; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL GALAXY G3; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Catalog Number GLX120408
Medical Device Problem Code Stretched (1601)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/18/2025
Type of Reportable Event Malfunction
Event or Problem Description
The healthcare professional reported that during a stent-assisted endovascular embolization, the physician filled the aneurysm with a 4mm x 8cm galaxy g3 xsft (glx120408 / 2022547s) coil, and it was found that the coil had unraveled / stretched.The physician retracted only the coil and replaced it; the replacement coil was delivered with the same original concomitant 150cm x 5cm prowler select plus microcatheter (606s255x / lot# unknown).During the delivery of the 4mm x 16mm enterprise 2 vascular reconstruction device (encr401612 / 9784479), the stent was impeded at the proximal end of the microcatheter and could not be further advanced.The physician retracted only the stent and delivered it again, but the same issue was encountered.The physician removed the stent and the microcatheter from the patient and replaced both devices to complete the procedure.There was no report of any negative patient impact.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, sex, gender, weight, race, and ethnicity were not provided.Section e.1: the initial reporter phone is not available / reported.Section h.3: the device is available to be returned for evaluation and testing.However, it has not been received to date.If the device returns, a device investigation will be performed.A review of manufacturing documentation associated with this lot (2022547s) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no internal actions related to device manufacture or inspection.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report that multiple attempts to obtain product for analysis were unsuccessful.If the product is returned or information provided at a later date, the file will be updated accordingly.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.A review of manufacturing documentation associated with this lot (2022547s) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no internal actions related to device manufacture or inspection.With the information available and without the complaint product available to be returned for analysis, the reported issue documented in the complaint cannot be confirmed through functional evaluation and analysis.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.Assignment of root cause for the event remains speculative and inconclusive, based on the limited information provided and without the return of the complaint sample; however, it is possible that clinical and procedural factors, including device manipulation / interaction may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.Updated sections: b.4, g.3, g.6.H.2, h.3, h.6, and h.11.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALAXY G3
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CERENOVUS INC. (FREMONT)
47709 fremont blvd
fremont CA 94538
Manufacturer Contact
henry boden
31 technology dr
irvine, CA 92618
949788-2078
MDR Report Key23979782
Report Number3008114965-2026-00050
Device Sequence Number19991036
Product Code KRD
UDI-Device Identifier10886704077732
UDI-Public(01)10886704077732(17)280731(10)2022547S
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K150319
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberGLX120408
Device Lot Number2022547S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/18/2025
Supplement Date Received by Manufacturer02/09/2026
Initial Report FDA Received Date01/06/2026
Supplement Report FDA Received Date02/09/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/24/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-