| Catalog Number |
7M2802 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
12/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
According to the complainant, the te fill needles broke off when using with allergan fill kit.Did not break on the hub but at the body of the needle.Needle was inside the patient.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).Seventy-two (72) unused samples in original packaging were returned for evaluation.Based on other confirmed complaints of the same nature, the reported defect is confirmed.A supplier corrective action request (scar) was issued, and the supplier conducted their investigation.The root cause for the dull needle complaints was identified as a supplier misinterpretation of the bevel angle specification, resulting in needles manufactured with an incorrect, thinner bevel more susceptible to damage during normal handling steps.Corrective actions were implemented to prevent future bevel damage and ensure the supplier adheres to the specified requirements.We will maintain this report for further references and continue to monitor other reports for similar occurrences.
|
| |
|