| Model Number |
2110 |
| Medical Device Problem Code |
Device Alarm System (1012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3: date of event is unknown; no information has been provided to date.One device was received for investigation.The reported issue was confirmed during functional testing when the reported issue was duplicated.The issue was traced to the downstream occlusion sensor, which was determined to be faulty.The product's history records were reviewed and there were no non-conformance's nor service-related issues that would have resulted in the reported complaint.The dso sensor was replaced and the device passed all testing.
|
| |
|
Event or Problem Description
|
|
It was stated that the device had a too low downstream occlusion.There was no reported patient involvement.
|
| |
|