• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL, INC. CADD-SOLIS AMBULATORY INFUSION PUMP; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL, INC. CADD-SOLIS AMBULATORY INFUSION PUMP; PUMP, INFUSION Back to Search Results
Model Number 2110
Medical Device Problem Code Device Alarm System (1012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3: date of event is unknown; no information has been provided to date.One device was received for investigation.The reported issue was confirmed during functional testing when the reported issue was duplicated.The issue was traced to the downstream occlusion sensor, which was determined to be faulty.The product's history records were reviewed and there were no non-conformance's nor service-related issues that would have resulted in the reported complaint.The dso sensor was replaced and the device passed all testing.
 
Event or Problem Description
It was stated that the device had a too low downstream occlusion.There was no reported patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD-SOLIS AMBULATORY INFUSION PUMP
Common Device Name
PUMP, INFUSION
Manufacturer (Section D)
ICU MEDICAL, INC.
3350 granada avenue north
oakdale MN 55128
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key23981061
Report Number3012307300-2026-00161
Device Sequence Number10608967
Product Code FRN
UDI-Device Identifier10610586041037
UDI-Public(01)10610586041037(11)160425
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K111275
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number2110
Device Catalogue Number21-2111-0400-51
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/10/2025
Initial Report FDA Received Date01/06/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/25/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-