| Model Number |
H938176 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1: initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported eight exactamix inlets had ¿movable and immovable black particles inside the packaging.¿ the particulate was observed prior to patient use; therefore, there was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added to g4, h6, h7, h9 and h11.Correction to h4: the lot was manufactured february, 2024.H11: the actual devices were not available; however, three (3) photographs of samples were provided for evaluation.Visual inspection was performed which identified two (2) photos with a dark particle embedded in the outside of the tubing, not in the fluid pathway.One (1) photo did not show any particulate matter.The reported condition was verified in two photos and not verified in one photo.The cause of the condition was determined to be contamination during the manufacturing or packaging process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|