Model Number MER3700
Medical Device Problem Codes Data Problem (3196); Application Program Freezes or Fails to Launch (4031)
Health Effect - Clinical Code Arrhythmia (1721)
Date of Event 12/19/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient experienced a ventricular tachycardia (vt) episode and manual delivery of anti-tachycardia pacing (atp) was attempted using the device.During the manual atp, the programmer froze and the device aborted the manual atp.The device continued with programmed therapy that the physician was able to inhibit using magnet response.The programmer unfroze, and manual atp was again attempted, however, an error message was noted.Additional information was requested.The patient was in stable condition.
 
Event or Problem Description
Additional information was received indicating the device was reprogrammed.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN¿ 2 PATIENT CARE SYSTEM (PCS) PROGRAMMER
Common Device Name
PACEMAKER/ICD/CRT NON-IMPLANTED COMPONENTS
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key23987137
Report Number2017865-2026-00457
Device Sequence Number17317209
Product Code OSR
UDI-Device Identifier05415067031174
UDI-Public(01)05415067031174(10)
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMER3700
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/20/2025
Supplement Date Received by Manufacturer01/09/2026
03/20/2026
Initial Report FDA Received Date01/07/2026
Supplement Report FDA Received Date01/30/2026
03/20/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-