| Catalog Number |
35251340S |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported: "the set screws from the gamma nail kit did not hold in the nail.Afterwards, a separately packaged set screw was opened, but this one also did not hold in the nail.Subsequently, a new nail was implanted, and the set screw that came with this nail worked as intended.".
|
| |
|
Additional Manufacturer Narrative
|
|
Once the investigation is completed, any additional information will be reported in a supplemental report.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed since the device was returned for evaluation and matches the alleged failure.The device inspection revealed the following: the received nail overall appeared to be free from any major damage externally.However, internally the threads from where the set screw is placed in were damaged right at the beginning.Since the threads of both the nail & the set screw were majorly deformed, the issue could be confirmed.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on available information, the root cause of issue is deemed to be user related.The cross threading between the screw is indicative of the fact that the screw wasn¿t aligned properly with the threads in the nail, leading impaired insertion.If any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
As reported: "the set screws from the gamma nail kit did not hold in the nail.Afterwards, a separately packaged set screw was opened, but this one also did not hold in the nail.Subsequently, a new nail was implanted, and the set screw that came with this nail worked as intended.".
|
| |
|