| Catalog Number |
GST45T |
| Medical Device Problem Codes |
Material Integrity Problem (2978); Ejection Problem (4009)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an unknown procedure, it was observed that the cartridge did not fit into the body smoothly and seems like the cartridge had been destroyed.There was no patient consequence nor surgical delay reported.No additional information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 1/7/2026.D4: batch #unk.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.D4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.H4: the device manufacture date is currently not available.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.No lot or batch number was provided; therefore, a device history could not be done.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent 3/2/2026.D4 batch #897c58.Corrected data d4: udi#.Investigation summary : the product was returned for evaluation.Visual inspection was conducted on the returned reload.Visual analysis of the returned sample determined that one gst45t reload was received with no apparent damage, with an opened package, and fully fired.No functional test could be performed due to the condition of the reload.The event described could not be confirmed as the reload was received fully fired with no damage noted and it was placed and removed with no issues noted.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 897c58, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|