Model Number M08575A
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/18/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The baxter technical support found the patient circuit kit needed to be replaced.The synclara cough system is intended for use on patients who are unable to cough or clear secretions effectively due to reduced peak cough expiratory flow or respiratory muscle weakness.The synclara cough system provides a noninvasive therapy designed for use by patients, caregivers, and healthcare providers.The system will simulate a cough to remove secretions in patients with a compromised peak cough flow.Possible adverse conditions listed in the device instructions for use include swallowing too much air resulting in the likelihood of vomiting and aspiration.The use of a face mask may result in discomfort, vomiting, or breathlessness in some patients.Close supervision throughout the treatment is necessary when this product is used by children or patients with physical limitations or impaired cognitive abilities.A search of the baxter maintenance records did not show baxter performed any preventative maintenance on this product.It is unknown if the customer performs preventative maintenance on their products.A replacement patient circuit kit was sent to the customer.Based on this information, no further actions are required at this time.Although there was no reported injury with this event, if the report of a damaged biofilter were to recur, it could potentially cause serious injury or death.Therefore baxter is reporting this event.
 
Event or Problem Description
On 18-dec-2025, a customer contacted technical service to report that synclara system service, non-ft sw, hc, na (product code m08575a, serial number (b)(6)), had biofilter cracked where it connects to the device.The biofilter was disposed of by husband.This occurred at home during patient use.There was no patient injury or death reported with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCLARA COUGH SYSTEM
Common Device Name
DEVICE, POSITIVE PRESSURE BREATHING, INTERMITTENT
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
HILL-ROM SINGAPORE
1 yishun avenue 7
singapore, north east 76892 3
SN   768923
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key23989775
Report Number3008145987-2026-00001
Device Sequence Number18772552
Product Code NHJ
UDI-Device Identifier00887761995062
UDI-Public(01)00887761995062(11)990127
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K192143
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberM08575A
Device Catalogue NumberM08575A
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/18/2025
Initial Report FDA Received Date01/07/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/27/1999
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-