|
SYNTHES GMBH ZERO-P VA CAGE LRG LORDOTIC H7 PEEK; INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, CERVICAL
|
Back to Search Results |
|
| Catalog Number |
04.647.227S |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that zero-p va cage lrg lordotic h7 peek and scrdriver t8 self-holding angl w/sleeve were involved in an intraoperative event during a spinal fusion procedure.During insertion of the inferior screw, two articulating drivers broke and the titanium portion of the cage detached.The pin at the articulating joint of the drivers sheared off.The cage was removed and replaced with a new one, and a straight shaft driver was used to complete the procedure.The event resulted in a 10-minute surgical delay, but the procedure was successfully completed without any adverse patient consequences.All device fragments were removed without additional intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.Investigation summary.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot.Part number: 04.647.227s-us.Lot number: 35047p3.Manufacturing site: mezzovico.Release to warehouse date: 10 sept 2024.Expiry date: 31 aug 2034.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.¿ if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|