(b)(4).Date sent: 1/7/2026.D4 batch #: y7010v.Investigation summary.The product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the device was returned with the clamp arm detached from the device and not returned.Due to the condition of the returned device, no all functional testing could be performed.The device was disassembled to inspect the blade and a crack was located at an area inside the tube proximal to the tissue pad.This is consistent with a side load being applied to the blade while active with the jaw closed.A manufacturing record evaluation was performed for the finished device lot/batch number, and no non-conformances were identified.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage.This in turn can result in activation issues such as failing the pre-run test with the generator and displaying an alert screen.These alert screens that can result are such as "remove instrument from patient" or ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damage blade can result in a broken blade.No conclusion could be reached as to how the blade crack issue occurred.Possible causes of the clamp arm damage are not closing the clamp when introducing or removing through the trocar; or possible entanglement in fibrous tissue.As part of the quality process all devices are manufactured, inspected, and released to approved specifications.
|