| Catalog Number |
BOPT5410 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Osteolysis (2377)
|
| Date of Event |
11/13/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(6).A4: weight unknown / not provided.D4: lot number unknown / not provided.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost non-existent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the implant at tooth site #3 was initially placed in 2017 with sinus bump and mineross.The implant failed first at uncover surgery and was removed due to peri-implantitis.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to relay corrected data and additional information.Correction: the device catalog number was updated from bnpt5410 to bopt5410.The lot number was updated from unknown to 2016070538.The following sections are being reported: b4: date of this report was updated.B5: additional information was received.D1: brand name was updated.D4: catalog number, lot number, unique device identifier (udi) number and expiration date were updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H4: device manufacture date was updated.H10: narrative/data was updated.
|
| |
|
Event or Problem Description
|
|
Additional information was received updating the catalog and lot number.
|
| |
|
Search Alerts/Recalls
|