DEPUY ORTHOPAEDICS INC US ATTUNE CR FEM TRIAL SZ 5 RT; FEMORAL TRIALS Back to Search Results
Catalog Number 254500705
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/23/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Broke 2 femoral sz5 right trials when placing the femoral component.2 replacements are required to be sent in to [hospital].Was surgery delayed due to the reported event? unknown, was procedure successfully completed? unknown, were fragments generated? unknown, if yes, were they removed easily without additional intervention? unknown, patient status/ outcome / consequences no, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: unknown, is the patient part of a clinical study unknown, (b)(4), device property of none, device in possession of none.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> according to the information received, "broke 2 femoral sz5 right trials when placing the femoral component.2 replacements are required to be sent in to [hospital]." the product was not returned to depuy synthes, however photos were provided for review.See attachments ¿img_7941.Jpeg¿, ¿img_7938.Jpeg¿, ¿img_7940.Jpeg¿ and ¿a-14474253 source data¿.The photo investigation revealed that '254500705, attune cr fem trial sz 5 rt¿ had a condyle broken.The fracture surface is consistent with overload due to a combination of heavy impaction and the trial fitting too tight over the posterior/anterior aspect of the resected femur bone forcing the curved features of the femoral trial to open.The combination of heavy impaction and possible discrepancies in the resection depth between the femur and the trial suggest unintended user error.Furthermore, per the attune surgical technique (dsus/jrc/0316/1437 rev.K) it is cautioned when extracting the trial, rocking the trial medio-laterally may cause condylar fracture.Such rocking should be avoided.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the attune cr fem trial sz 5 rt would contribute to the complained device issue.Based on the investigation findings, potential cause can be attributed to unintended use error and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot
=
> the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CR FEM TRIAL SZ 5 RT
Common Device Name
FEMORAL TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23993981
Report Number1818910-2026-00324
Device Sequence Number10184642
Product Code HWT
UDI-Device Identifier10603295133773
UDI-Public(01)10603295133773(10)MVMBSC110
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254500705
Device Lot NumberMVMBSC110
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/22/2025
Supplement Date Received by Manufacturer03/09/2026
Initial Report FDA Received Date01/07/2026
Supplement Report FDA Received Date03/10/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/01/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-