Model Number 1103
Medical Device Problem Code Pumping Problem (3016)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/30/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation of this event is pending and a supplemental report will be sent upon its completion.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the ventricular assist device (vad) had a motor start event with starting parameters outside of the typical range noted in log files data following a routine controller exchange.The vad remains in use.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
A supplemental report is being submitted for investigation summary.Product event summary: the ventricular assist device (vad) (b)(6) was not returned for evaluation.No performance allegations were made against the device; therefore, the purpose of this investigation is solely to address the finding identified during analysis of the controller log files.A review of the manufacturing documentation was not performed as the reported event and analysis are not related to a manufacturing or servicing issue.Log file analysis revealed a motor start event recorded on 27/jan/2022 at 17:08:03, in which the motor start parameters were atypical.As a result, the atypical motor start parameters finding was confirmed.Based on risk documentation, the most likely root cause of atypical motor start parameters may be attributed, but not limited, to outer shroud contact that creates more friction at the housing to impeller interface during pump startup.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
A supplemental report is being submitted for a correction to the event date.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer (Section G)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key23994732
Report Number3007042319-2026-00021
Device Sequence Number21378635
Product Code DSQ
UDI-Device Identifier00888707003261
UDI-Public(01)00888707003261(17)200930
Combination Product (Y/N)N
Initial Reporter StateDC
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/30/2020
Device Model Number1103
Device Catalogue Number1103
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/30/2025
Supplement Date Received by ManufacturerNot provided
01/27/2026
Initial Report FDA Received Date01/07/2026
Supplement Report FDA Received Date01/21/2026
02/02/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/26/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age51 YR
Patient SexFemale
-
-