| Model Number |
CA461HS |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer was contacted in reference to the voluntary field safety notice/recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.A remstar pro c-flex+ device was returned to third party service center for evaluation, there was no report of serious or permanent harm or injury.There was no report of medical intervention.During evaluation no foam particle were noted in the air path, yet foam degradation was noted.The device sound abatement foam needs replaced to address the issue.In addition, error code: 66 (e-66: err_stuck_encoder_b), error code: 65 (e-65: err_stuck_encoder_a), error code: 62 (e-62: err_stuck_ramp_key) and error code: 63 (e-63: err_stuck_knob_key).Dust was found as contamination.The device as scrapped.
|
| |
|
Additional Manufacturer Narrative
|
|
In previously submitted report, operator of device in box d and occupation in box e, report source in box g, reason device not evaluated and health impact code were not mentioned correctly in box h was incorrect.In this report, operator of device in box d and occupation in box e, report source in box g, reason device not evaluated and health impact code in box h has been updated/corrected.
|
| |
|