• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON & CO (FRANKLIN LAKES) BD VACUTAINER® SERUM; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON & CO (FRANKLIN LAKES) BD VACUTAINER® SERUM; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION Back to Search Results
Catalog Number 367814
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/12/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported while using bd vacutainer® serum, tube push off occurred with 20 tubes.No adverse impact was reported regarding the patient or user.
 
Event or Problem Description
It was reported while using bd vacutainer® serum, tube push off occurred with 20 tubes.No adverse impact was reported regarding the patient or user.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information there were multiple 510k numbers reported to be involved.The information is as follows g.4.Pma / 510(k)# k960250.Investigation summary: bd had not received any samples or photos for investigation.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: tube push off.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SERUM
Common Device Name
TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION
Manufacturer (Section D)
BECTON DICKINSON & CO (FRANKLIN LAKES)
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
anand singh
1 becton drive
franklin lakes, NJ 07417
2018475186
MDR Report Key23995951
Report Number1024879-2025-02380
Device Sequence Number10399112
Product Code JKA
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
BK050036
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number367814
Device Lot Number5105267
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/12/2025
Supplement Date Received by Manufacturer03/17/2026
Not provided
Initial Report FDA Received Date01/07/2026
Supplement Report FDA Received Date04/08/2026
09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/15/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-