| Model Number |
420194-13 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was a broken cable with a da vinci product.The customer reported event has no report of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The instrument involved in this complaint has not been received and/or tested by failure analysis as of the date of this report.If the product is received and evaluated, a supplemental mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis, and the complaint was confirmed.The monopolar curved scissors (mcs) instrument was analyzed and found to have a broken grip cable at the proximal end within the housing.The grip cable was fully broken at the input shaft.There was no discoloration, corrosion, or contamination on the cable the suggest improper reprocessing as a contributing factor.Further inspection revealed no signs of missing parts and found no abnormally sharp or damaged components that could have contributed to the cable breaking.The probable root cause of cable breakage is attributed to damage to the cable through external collisions, during use, or during reprocessing.Correction(s): d4.Lot number was updated to: k10240125 682.
|
| |
|
Event or Problem Description
|
|
Refer to h11 for follow-up information.
|
| |
|