| Catalog Number |
0500325E |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
|
| |
|
Event or Problem Description
|
|
A user facility office manager reported that a visible external dialyzer blood leak occurred thirty minutes after the initiation of the patient¿s hemodialysis (hd) treatment.The machine, a fresenius 2008t machine, did not alarm, but is not expected to for an external blood leak.A crack was noted on the dialyzer.Fresenius bloodlines were not in use.The patient¿s blood was completely returned following the incident.The patient¿s estimated blood loss (ebl) is unknown.There was no patient serious injury or medical intervention required as a result of this event.The patient was restarted on the same machine and treatment completed successfully with new supplies.The complaint device is not available to be returned to the manufacturer for evaluation.
|
| |
|
Event or Problem Description
|
|
A user facility office manager reported that a visible external dialyzer blood leak occurred thirty minutes after the initiation of the patient¿s hemodialysis (hd) treatment.The machine, a fresenius 2008t machine, did not alarm, but is not expected to for an external blood leak.A crack was noted on the dialyzer.Fresenius bloodlines were not in use.The patient¿s blood was completely returned following the incident.The patient¿s estimated blood loss (ebl) is unknown.There was no patient serious injury or medical intervention required as a result of this event.The patient was restarted on the same machine and treatment completed successfully with new supplies.The complaint device is not available to be returned to the manufacturer for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: the reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.Continuous improvement is of the utmost importance to fresenius medical care as we strive to provide dialysis products of the highest quality to our patients.Reports of leaking product are investigated both individually as complaints, as well as via the nc/capa program, in order to assess and improve our products and processes.Capas for vision systems and blood leak reduction are recent examples of leak related investigations directed at an overall reduction in dialyzer leaks.
|
| |
|
Search Alerts/Recalls
|