| Model Number |
LF-TP |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation found the cement glue on the light guide (lg) glass/lens cover is missing.Based on the results of the investigation, it is likely that the following led to the malfunction: cause traced to component failure.A definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the tracheal intubation fiberscope exhibited the cement glue on the light guide (lg) glass/lens cover is missing.There was no patient involvement.
|
| |
|