• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. 12.5MM ENTRY REAMER; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. 12.5MM ENTRY REAMER; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Catalog Number 71631116
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/12/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an internal fixation surgery, the 12.5mm entry reamer was dull and wouldn't cut, the ruler had a broken internal plastic ring which caused it to give a fixed reading, the shaft of a medium hexdriver wouldn't fit into the sleeve, it got stuck, and wouldn't rotate to lock the screw to the sleeve, and it was so tight that it prevented the screwdriver shaft from being easily inserted or removed.The procedure was resumed, after a significant delay, using the same devices.No further complications were reported.
 
Additional Manufacturer Narrative
H11: internal complaint reference (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.The following are the part numbers reported: 71631116, 71674079, 71631066.However, it is unknown which of the devices contributed with the delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12.5MM ENTRY REAMER
Common Device Name
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key24003153
Report Number1020279-2026-00030
Device Sequence Number13317520
Product Code JDG
UDI-Device Identifier03596010419583
UDI-Public3596010419583
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71631116
Device Lot Number17JM04527
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/12/2025
Initial Report FDA Received Date01/08/2026
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MEDIUM HEXDRIVER PN: 71631066 LN: 22MSY0026; RULER PN: 71674079 LN: 20GCT0048
Patient Age30 YR
Patient SexMale
Patient Weight100 KG
-
-